The firewall and SE Linux are disabled by default (otherwise the following ports must be open for UDP: 52010, 52011,
52012). The following firewall port must be open if you want to connect to Camserver with a TCP/IP socket connection from
the outside: 41234
4.2.3. Samba Share
There is a Samba share configured on the DCU. The Samba service is enabled by default and running after the system
boots.
The storage directory for the images, /home/det/p2_ det/images, which is a symbolic link to /home/det/images, can be
accessed from a Samba client. You need the same user name and password as for the normal console or ssh login.
From a Linux or Mac OSX terminal window issue the following command:
smbclient
server-ip-or-hostname/images -U det
To browse the Samba share from Windows, type
in the Windows Explorer address bar.
4.3. Thermal Stabilization Unit
A thermal stabilization unit is required for the operation of the PILATUS3 S 6M detector system. The hoses and the detec-
tor are equipped with self-sealing valves to avoid dripping when connecting or disconnecting the tubes. There is no fixed
limitation on the length of the tubing, but it should be kept as short as possible to ensure the best flow.
Table 4.6:
In-air operating conditions
Condition
Definition
Operating temperature
The thermal stabilization unit has to be set to a temperature
of 23
◦
C for normal operation.
Maximum operating pressure
3 bar
Coolant
66 %
distilled
water
and
34 %
ethylene
glycol.
Danger
#2
Ethylene glycol can be seriously harmful to your
health or fatal if handled incorrectly. Consider
the packaging and safefty instructions provided
by your local supplier.
Information
#6
Before operating the thermal stabilization unit, please read the User Manual of the thermal stabilization unit.
Please consider the following points for installation and usage of the thermal stabilization unit:
• When connecting or disconnecting the cooling hoses, turn off the detector and the thermal stabilization unit.
• When operating the detector, the thermal stabilization unit must always be turned on and the pump has to be activated
(see user documentation of thermal stabilization unit).
• Use opaque hoses to avoid the growth of algae.
• Do not set the temperature of the thermal stabilization unit below the recommended operating temperature. Condens-
ing moisture can develop and damage the detector.
PILATUS3 S 6M Technical Specifications v1.1.1
14 | 25