4.1.4. Status LEDs
Table 4.1:
The meaning of the status LEDs on the detector back plane
LED
Description
EN
Orange to indicate the detector is in counting mode, but is other-
wise off.
TEMP
Red at power-on, when the detector is not initialized, or if the de-
tector temperature or humidity is out of the limits. Turns green
on detector initialization if temperature and humidity are within the
allowed limits.
POWER
Green when the power supply is functioning. Red is an indication
of power failure or if the detector temperature or humidity is out of
the operating limits.
4.1.5. Connectors and Connecting Cables/Pipes
Table 4.2:
Electric Connectors and Connecting Cables
Connector
Description
DATA
2 x RJ45 Ethernet connector
DATA 1
→
detector control unit port 1
DATA 2
→
detector control unit port 2
Use Cat 6A S/FTP Ethernet cable(s).
Caution
#5
There must be a 2 x 10GBase-T Ethernet point-to-
point connection between detector and detector con-
trol unit.
POWER
AC power connector (see table 3.3)
EXT IN
External trigger input (see table 3.3)
Use a Lemo
®
Type 00 (NIM/CAMAC) cable.
EN OUT
Output signal 5 V (max. current 100 mA) (see table 3.3)
High when counting is enabled.
Use a Lemo
®
Type 00 (NIM/CAMAC) cable.
Functional ground
Information
#4
Although the detector might be already grounded via
the mounting bolts, the detector should be grounded
additionally via the functional ground connector at the
back to establish a defined grounding.
PILATUS3 S 6M Technical Specifications v1.1.1
11 | 25