3.4. Vacuum Conditions for Detectors with Optional Vacuum Compatibility
DECTRIS
®
detectors can be provided for vacuum operation. The typical reachable vacuum is 10
−3
mbar (read-out elec-
tronics in vacuum) or 10
-6
mbar (only detector head in vacuum). The out-gassing rate and content are not specified. The
vacuum compatibility option guarantees that the detector will operate in the mentioned vacuum range. Detectors ordered
with a vacuum compatibility option have been tested in vacuum.
To see if a detector is vacuum compatible, check the conditions mentioned in section 2.1
Warning
#3
Only systems with a vacuum compatibility option are allowed to be operated in vacuum. Please contact
[email protected] for information regarding vacuum compatibility upgrade.
For in-vacuum operation of the detector following conditions must be fulfilled:
Table 3.8:
In-Vacuum Operating Conditions
In-Vacuum Condition
Definition
Pressure during operation
atmospheric pressure or less than 0.01 mbar
1 bar
10
-
² mbar
1 Pa
10⁵ Pa
...
Detector mounting plate temperature during
operation
+10
◦
C to 25
◦
C
Thermal stabilization unit set temperature in vacuum
10
◦
C
Chamber temperature during ”bake-out” (detector
unpowered)
max. +60
◦
C (for temperatures > 40
◦
C make sure the thermal
stabilization unit is set to +40
◦
C and running)
PILATUS3 R 100K-M Technical Specifications v1.2.2
8 | 34