Page: 30 of 102
6.2
Construction OMD-2008 EVA
The OMD-2008 EVA consists of a Computer Unit, and a Measuring Cell, which is equipped
with an EV-Valve and Automatic Cell Cleaning Device. Additionally, the components of the
OMD-2008 EVA are mounted onto three separate supports.
The Automatic Cell Cleaning Device is electrically inserted between the Computer Unit
and the Measuring Cell. The cable length is limited to typical about 1 m. The cable is
connected with the Measuring Cell as well as with the Automatic Cell Cleaning Device
consecutively.
The Automatic Cell Cleaning Device additionally needs a sufficient supply of dry, clean
instrumentation air.
Figure 10. Construction OMD-20008 EVA
1. Computer Unit
6.
Automatic Cell Cleaning
Device
10. Autoclean Controller
2. Measuring Cell
7. Desiccator Cap
11. Air Regulator
3. Electric Valve
8. Communication Cable
(Measuring Cell)
12. Air Flow Adjuster
4.
EV Valve
Connection Cable
9. Communication Cable
(Automatic Cell Cleaning
Device)
13. Air Supply for Regulator
5. EV Support
14. Air Supply for Cylinder
Содержание OMD-2008
Страница 28: ...Page 27 of 102 5 5 Cabling OMD 2008 EV Figure 9 Bottom View OMD 2008 EV ...
Страница 84: ...Page 83 of 102 ...
Страница 85: ...Page 84 of 102 ...
Страница 86: ...Page 85 of 102 ...
Страница 87: ...Page 86 of 102 ...
Страница 88: ...Page 87 of 102 ...
Страница 89: ...Page 88 of 102 ...
Страница 90: ...Page 89 of 102 ...
Страница 91: ...Page 90 of 102 ...
Страница 92: ...Page 91 of 102 ...
Страница 93: ...Page 92 of 102 ...
Страница 94: ...Page 93 of 102 ...
Страница 95: ...Page 94 of 102 ...
Страница 99: ...Page 98 of 102 Type Approved by ...