![Decagon Devices EC-5 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/decagon-devices/ec-5/ec-5_user-manual_2461469020.webp)
EC-5 User’s Manual
4. Collecting Data
16
Rockwool
The EC-5 was calibrated in Grodan Master rockwool
with solution electrical conductivities of 0.2, 1.0, 1.5,
2.0, and 4.5 dS/m. Volumetric water content can be cal-
culated using
ø = 6.28 * 10
-7
*
RAW
2
+ 1.37 x 10
-4
*
RAW
- 0.183(5)
for a Decagon data logger or
ø = 2.63 * 10
-6
* mV
2
+ 5.07 x 10
-4
*
mV - 0.0394(6)
for a data logger with 2500 mV excitation.
NOTE
:
These calibration constants only apply to
2500mV excitations; use of these numbers with any other
excitation voltage will result in erroneous readings!
Troubleshooting
If you encounter problems with the EC-5, they most
likely will manifest themselves in the form of incorrect
or erroneous readings. Before contacting Decagon about
the sensor, do the following:
• Check to make sure the connections to the data
logger are both correct and secure.
• Ensure that your data logger’s batteries are not
dead or weakened.
If you encounter problems that aren’t due to the data
logger, please contact Decagon at
(509) 332-5600