background image

     

 

 

10

 / 

11

 

 

WWW.DDTOP.COM 

 

 

Figure5-1                                  Figure5-2                                Figure5-3                    Figure5-4       

 
3.  All  terminals  shall  be  in  good  contact  with  each  other,  and  the  nut  of  outgoing  wire  shall  be 
properly pressed to prevent the lead from loosening. 
4.  Make  sure  that  excess  lines do  not  interfere with the  movement  of  the  switch  cover  or  switch 
mechanism. 

 

7.3 Debugging Operation Process 

1. The instrument should not be subjected to strong vibration and impact, especially on the float 
level switch should not be strongly tension pressure, so as to avoid damage to the spring. 
2.The  instrument  may  not  participate  in  impactful  process  tests  such  as  air  sweeps  performed 
before the device is put into production. 
3.During  commissioning,  if  water  is  used  as  the  medium  (not  actually  measured  medium),  the 
following two situations will occur:  ρ  medium <ρ  water and  ρ  medium>ρ  water. If  ρ  medium <
ρ  water, it can be debugged with water. If  ρ  medium>ρ  water, it can't be debugged with water. It 
can only be debugged with actual measuring medium. 
4. For  ρ  medium <ρ  water When calibrating with water, the alarm position will be lower than the 
alarm position required by the actual measurement medium. The difference will be different due to 
different working conditions and cannot be calculated by formula. 

8. Precautions 

8.1 The design density should be the same with the design density. If the density changes, please 
contact the manufacturer to confirm whether it can continue to be used.                                                   
8.2    During the installing and uninstalling process, the torque tube should be protected as much as 
possible to prevent damage.     

8.3 The displacer level controller is not suitable for high vibration working condition. 

8.4 Medium with viscosity higher than 500mPa/s is not suitable for use with the displacer level 

controller. 

8.5 Medium with ferromagnetic particles are not suitable for use with displacer level controller. 

8.6 After wiring, tighten the power supply cover to prevent the instrument from getting wet. 

Содержание UTK

Страница 1: ...1 11 WWW DDTOP COM UTK DISPLACER LEVEL SWITCH Operation Manual UTK DT JS 1012 2018 A...

Страница 2: ...ut prior notice Quality Assurance Dandong Top Electronics Instrument Group Co Ltd guarantees that all glass plate level gauge have no defects in materials and manufacturing processes within one year f...

Страница 3: ...3 Technical Characteristics 7 3 1 Main Performance 7 3 2 Main Parameters 7 4 Dimensional Outline Drawing Figure 2 7 5 Unpacking and Inspection 8 5 1 Unpacking Inspection Notice 8 Figure 4 Sample Name...

Страница 4: ...ment in intrinsically safe or non flammable site operating procedures 1 2 Process leaks can cause serious injury or death Care should be taken to lift the transmitter If the process seal is damaged th...

Страница 5: ...0 Spring assembly 11 Displacer 12 Wire rope 2 2 Operating Principle 2 21 Classification of Displacer Level Switch The Displacer Level Switch is divided into ball float type and displacer type based on...

Страница 6: ...an be controlled and given an alarm 2 1 3 Displacer Type Operating Principle With the displacer as the sensing element the displacer hangs on a spring through the suspension rod When the level is rais...

Страница 7: ...rmance 3 1 1 Has passed the national explosion proof certification certification mark Intrinsically safe Ex ia CT1 T5 T6 Explosion isolation Ex d CT1 T5 T6 3 1 2 Product Standard Q AMM 004 Displacer L...

Страница 8: ...Figure 4 is consistent with the information in the supply list 5 1 2 Check the quantity and material of each part against the packing list 5 1 3 Check the appearance of the instrument for defects dama...

Страница 9: ...ation for Commissioning 7 1 1 Debugging tools Multimeter 7 2 Electrical Wiring 7 2 1 Wiring Cable Preparation The wiring cable shall be outer diameter 10mm five core rubber sleeve soft wire or rubber...

Страница 10: ...with water If medium water it can t be debugged with water It can only be debugged with actual measuring medium 4 For medium water When calibrating with water the alarm position will be lower than the...

Страница 11: ...ctual alarm position and specification 1 Process condition is not consistent with design condition The actual density or alarm setting value is inconsistent with the design value 2 On site heat tracin...

Отзывы: