CAL3K-A INSTALLATION GUIDE V1.1 |
(Press repeatedly)
You will see the command (Line#3) and a short explanation (Line#4). If you wish to operate the indicated command,
press
To see the FULL list of commands:
Type
(PASSWORD will be displayed on screen), now press
Type
(
DDS3K or your current password
) and press
If you hear a loud beep after pressing enter, you’ve entered the wrong password.
Function keys: They serve as a shortcut to the keyboard commands. The following function keys are used:
F1
Mass entry
(F1,Enter,0.xxxx,Enter)
F2
Sample Identification
(F2,Enter,mysampleId,Enter)
F3
Group Identification
(F3,Enter,MyGroupId, Enter)
F4
Spike Mass
(F4,Enter,0.xxxx,Enter)
F5
Balance Mass
(F5,Enter)
F6
Open Lid
(F6,Enter,Y,Enter)
F7
Vessel User Test
(F7,Enter)
F8
Terminate User Test
(F8)
F9
CAL3K-A_AP User Test
(F9,Enter)
F10
Help
(F10) Read display!
MASS ENTRY
The mass is a weight in the range 0.2 to 1.5 Gramm. The limits can be set via a PC to suit your
operation. Once it is entered it is ‘LOCKED’ in to the present operation until it is ignited. The only way
to ‘ERASE’ the mass or ‘UNLOCK’ it is as follows: