SERIES 5 SIDE DELIVERY SYSTEM 

Instructions For Use

Содержание DW5050

Страница 1: ...SERIES 5 SIDE DELIVERY SYSTEM Instructions For Use...

Страница 2: ...2 OVERVIEW 2 SERIES 5 SIDE DELIVERY SYSTEM Instructions For Use SIDE DELIVERY SYSTEM DW5050 DW5550...

Страница 3: ...TENTS 2 3 4 5 6 7 8 9 10 14 15 18 WARNING Failure to carefully follow the described procedure may result in damage to the equipment the operatory and or the operator NOTE Take note of additional impor...

Страница 4: ...ws as a regular maintenance item to ensure proper positioning of the device IEC Medical Classification 2A TYPE BF Power optics Splash Protection IPX0 TECHNICAL DOCUMENTATION The manufacturer will make...

Страница 5: ...uted without increase risk of shock or fire Use authorized replacement parts only Power cords must be installed by qualified personnel Ensure all service loops strain reliefs and cord guards are in pl...

Страница 6: ...YSTEM Instructions For Use Unit Keypad Function List Key Description Initiate Function Cancel Function Press key to intiate function Press any key to cancel function Release key to cancel function Pre...

Страница 7: ...7 OPERATION CONTROL HEAD 7 SERIES 5 SIDE DELIVERY SYSTEM Instructions For Use CONTROLS...

Страница 8: ...ieces Rotate counterclockwise to increase flow clockwise to decrease flow HANDPIECE FLOW CONTROLS Located at the front of the control head these individual control valves adjust the amount of water co...

Страница 9: ...e is observed turn the Water Bottle On Off switch OFF to release all pressure before tightening the bottle to stop leakage 2 Select either city water or bottled water supply source as desired using th...

Страница 10: ...rclockwise to increase pressure DRIVE AIR FLOW ADJUSTMENT DRIVE AIR PRESSURE GAUGE UTILITY CENTER The Utility Center comes factory preset at 40 psi for Water pressure and 80 psi for Air pressure All r...

Страница 11: ...inge to prevent splashing and to achieve a desirable mist pattern The adjusting screws are located under the unit head directly below the syringe control block SYRINGE CONTROL BLOCK ADJUSTMENT SCREWS...

Страница 12: ...nuckle B 5 When the unit head is level retighten the set screw in the stop block to secure the adjustment A HEAD LEVELING FRONT FLEX ARM AND HEAD ROTATION Rotational tension is preset at the factory h...

Страница 13: ...knuckle clockwise with a 3 16 hex key B Retighten the knuckle block set screw A Replace the end caps TO DECREASE SPRING TENSION 1 Remove the end cap at the front of the flex arm and loosen the set scr...

Страница 14: ...es at their recommended strengths can be corrosive at higher than recommended dilution ratios CHEMICAL DISINFECTION CAUTION These disinfectants will harm the surface finish of dental equipment and are...

Страница 15: ...rovide additional recommendations regarding this procedure including information on frequency and duration of flushing and the use of antibacterial solutions in the self contained water system MAINTEN...

Страница 16: ...stem again until all the cleaning solution is used up 4 Purge the unit with air a Hold the handpiece tubing and syringe over a container Turn the Master Switch to the On position wait a few moments th...

Страница 17: ...cleaning procedures must be performed in a consistent regular manner Failure to do so could result in contaminated water lines and a lower water quality than what is acceptable for patient use Follow...

Страница 18: ...ld be worn to reduce the possibility of damage ACCESSORY USE Using accessory devices not specified by DCI for use with their equipment may result in an increase of electromag netic emissions and or a...

Страница 19: ...DCI International LLC 305 N Springbrook Road Newberg OR 97132 USA PN 50196 Rev A 2021 08 23 www dcionline com 800 624 2793...