USER GUIDE SALINITY SENSOR WITH MODBUS RTU OUTPUT MBRTU-SAL

SALINITY SENSOR WITH
MODBUS RTU OUTPUT
MBRTU-SAL

Содержание MBRTU-SAL

Страница 1: ...USER GUIDE SALINITY SENSOR WITH MODBUS RTU OUTPUT MBRTU SAL SALINITY SENSOR WITH MODBUS RTU OUTPUT MBRTU SAL...

Страница 2: ...nitoring of the environment Application scope Marine industrial waste water pharmaceutical biotechnology industrial manufacturing and other online whole process monitoring Features 1 Digital sensor di...

Страница 3: ...ironment 0 65 0 6MPa Calibration Method Two point calibration Respond Time 10 seconds T90 Temperature Compensation Automatic temperature compensation PT1000 Power Supply 12 24VDC 10 10mA Size Diameter...

Страница 4: ...ment at the place where the water flow is turbulent and reduce the influence of water bubbles on the measurement Keep the measuring probe 2cm away from the bottom The probe of the sensor is fouling or...

Страница 5: ...1 Salinity value R PSU value x10 for example the salinity value of 12 1psu is displayed as 121 with 1 decimal place by default 1 2 bytes unsigned short 0x1000 Temperature calibration R W Temperature c...

Страница 6: ...t send 1 2 3 4 5 6 7 8 ADR 03H Start register high byte Start register low byte Register number high byte Number of registers low byte CRC low byte CRC high byte The first byte ADR slave address code...

Страница 7: ...w byte address Data high byte Data low byte CRC code Low byte CRC code High byte When the slave receives correctly the slave sends back 1 2 3 4 5 6 7 8 ADR 06 Register high byte address Register low b...

Страница 8: ...rt after the change The baud rate saves the upper computer setting after power failure Baud rate support 4800 9600 19200 The baud rate of integer value allocation is as follows Integer Baud rate 0 480...

Страница 9: ...decimal place reserved Register returns hexadecimal unsigned integer data salinity value integer 10 1 decimal place reserved d Humidity measurement command Address 0x0107 40108 Number of registers 1 F...

Страница 10: ...ration in air Send command 01 06 10 01 00 00 DC CA Respond 01 06 10 01 00 00 DC CA c Salinity slope calibration Address 0x1003 41004 Number of registers 1 Function code 0x06 Calibration example calibr...

Страница 11: ...Manufacturer Daviteq Technologies Inc No 11 Street 2G Nam Hung Vuong Res An Lac Ward Binh Tan Dist Ho Chi Minh City Vietnam Tel 84 28 6268 2523 4 ext 122 Email info daviteq com www daviteq com...