Datex-Ohmeda S/5 monitors
58
Document No. 8005571-1
Max. press
indicates the pressure at which the safety valve opens and is normally 300
±
15 mmHg
for adult and 150 mmHg
±
15 mmHg for infant.
2 s after stop
indicates the pressure at 2 seconds
after the pump has stopped and is normally > 270 mmHg for adult and > 130 mmHg for infant. If
the value is less, check leakage by the active leak test.
5.1.5 NIBP pulse valve menu
Start Test
Start test
is for starting and
Stop test
is for stopping the test.
Set Valve
Set Valve
lets you adjust the opening of the pulse valve.
Pulse Valve Data
See NIBP Service menu in chapter
for information on general items Pressure, Zero, Protect
handle, Calibr. prot., +15 V power, AD0
to
AD7
as well as
Timeouts etc.
Pulse Valve Checking
Wrap an adult cuff around a pipe and connect the cuff to the module. Select the
Start test
and
push the ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The
module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and
displays it on the screen. The test can be manually stopped by selecting
Stop test
.
The valve can be adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First select
Set Valve and push the ComWheel. See the pulse valve value and adjust it by turning the
ComWheel. Then push the ComWheel to confirm the value.
The ‘
Interval 240 mmHg -> 50 mmHg
’ time should be less than 60 seconds when the valve is
‘150’ and less than 10 when fully opened (255). When fully closed (0), the system should be
airtight and the pressure does not drop. Depending on an individual, the pulse valve may remain
closed up to approx. value 100.
If the measured time deviates much from those above, then the pulse valve or its tubes are faulty.
Содержание S/5 PRESTN
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Страница 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Страница 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Страница 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Страница 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...