S/5 M-PRESTN Modules
43
Document No. 8005571-1
4 TROUBLESHOOTING
4.1 Troubleshooting charts
See also the
User’s Reference Manual
for more troubleshooting procedures.
4.1.1 NIBP
TROUBLE CAUSE
TREATMENT
No NIBP value displayed
NIBP not selected on screen.
Check monitor setup.
NIBP menu fading
No M-PRESTN module, module not
properly connected, or NIBP and
PRESTN module connected at the
same time.
Plug in the module.
Artifacts-message
Unsuccessful measurement due to
patient movement, shivering,
external artifact or weak signal.
Weak pulsation-message
Weak or unstable oscillation pulses
due to:
•
artifacts
•
weak pulse pressure due to
arrhythmias
•
improper cuff position or
attachment
•
too few pulses detected
•
weak or unusual blood
circulation
•
obese subject
Check patient condition and retry.
Check any leaks and retry.
Use proper size of cuff. Check
attachment.
Call service
Error X-message
NIBP hardware error.
X = error number.
See the description of the error message
code in
solutions listed in the next chapter.
Содержание S/5 PRESTN
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Страница 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Страница 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Страница 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Страница 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...