![Datex-Ohmeda S/5 M-P Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-p/s-5-m-p_technical-reference-manual-slot_513846030.webp)
Datex-Ohmeda S/5 monitors
Document No. 8001013-1
26
values do not indicate a failure, but the continuous counting (more than 50 per second) during the
normal operation indicates either serial communication failure or module not in place.
RAM
indicates the state of the RAM memory.
ROM
indicates whether the checksum in the EPROM
is in accordance with the one the software has calculated.
EEPROM
indicates if the values stored
in the permanent memory are valid. The state is either
OK, Fail
or
?
(module not in place or a
communication error).
5.2
P/PT calibration menu
Protection
Protection for the configuration and temperature calibrations can be set ON and OFF only when
protect button at the bottom of the module is pressed.
Set Config
The module configuration should be set according to the module type. The setting is possible only
when the protection is set OFF. The available selections are BP or PT.
The configuration setting should be checked if the PT board is replaced.
Calibrate T3 and Calibrate T4
The functions are for calibrating the temperature channels T3 and T4. The calibrations are possible
only when the protection is set OFF. The temperature calibration requires accurate test plugs of
value 25 °C and 45 °C.
Calibration:
1.
Select Calibrate T3 / Calibrate T4
2.
Insert the test plug 25 °C into the T3 / T4 connector
3.
Press the ComWheel
4.
Insert the test plug 45 °C into the T3 / T4 connector
5.
Press the ComWheel
Calibrate P3
The function is for calibrating the invasive blood pressure channel P3.
The calibration requires a pressure transducer (with an appropriate cable) and a pressure
manometer.