![Datex-Ohmeda S/5 M-P Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-p/s-5-m-p_technical-reference-manual-slot_513846022.webp)
Datex-Ohmeda S/5 monitors
Document No. 8001013-1
18
All modules
16.
Perform electrical safety check and leakage current test.
?
17.
Check that the module functions normally after the performed electrical safety check.
?
18.
Clean the module with suitable detergent.
?
•
Fill in all necessary documents.
3.3
Disassembly and reassembly
Disassemble the M-P and M-PT modules in the following way. See the exploded view of the
module.
1.
Remove the two screws from the back of the module.
2.
Pull the module box slowly backwards and detach it from main body. Be careful with loose
latch and spring pin for locking.
3.
To detach the PT board. The board can be removed by detaching two screws on the folio
side of the board near the front panel and disconnecting the two ribbon cables coming from
the front panel.
To reassemble the module, reverse the order of the disassembly steps.
CAUTION
When reassembling the module, make sure that all cables are connected properly.