Pressure Module, M-P and Pressure Temp Module, M-PT
Document No. 800 1013-1
13
3
SERVICE PROCEDURES
3.1
General service information
Field service of the M-P and M-PT modules is limited to replacing faulty circuit boards or
mechanical parts. The circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full
type designation, and a detailed description of the fault.
CAUTION
Only trained personnel with appropriate equipment should perform the tests and
repairs outlined in this section. Unauthorized service may void warranty of the unit.
3.2
Service check
These instructions include complete procedures for a service check. The service check is
recommended to be performed after any service repair. However, the service check procedures can
also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (
Appendix A
) which should be filled in when performing the
procedures.
The mark
?
in the instructions means that the check list should be signed after performing the
procedure.
The procedures are designed for monitors with S/5 monitor software of revision 01. However, most
of the procedures also apply to monitors, which contain some other monitor software
type/revision.
3.2.1
Recommended tools
Tool
Order No.
Notes
M-NE(12)STPR/M-ESTPR/M-ESTP
Not M-ESTP Rev. 01, M-EST Rev. 00 or M-ETP
Rev. 00
Patient simulator
Pressure manometer
Temperature test set
884515
InvBP transducer
Oscilloscope
Screwdriver