Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
44
Rate of the side flow is checked by blocking the side flow after the water trap and measuring the
flow rate as above. The rate should decrease by 10 to 27 ml/min.
Measurement flow and reference flow of the oxygen measuring unit are checked as follows:
1.
Connect the flowmeter behind the flow cassette (no. 2) ahead of the oxygen measuring unit
REF inlet. The flowmeter should show between 25 and 42 ml/min. The flow rate is adjusted
by changing the cassette.
2.
Connect the flowmeter between the oxygen measuring unit IN inlet and the tube which is
connected to it. The flow rate should be between 18 and 25 ml/min
larger
than the REF
flow. This is adjusted by changing the flow cassettes (no. 4 and 5) which are located
between the IN and OUT inlets.
3.
Flow rate of CO
2
absorber is measured by connecting the flowmeter to the unoccupied
connector of the flow cassette (no. 1). Make sure that the monitor is in normal situation
(APNEA text on the screen). The flow rate should be zero. When the gas zeroing takes place,
the rate should be between 180 and 220 ml/min. The gas zeroing can be simulated in the
ACX Service Menu manually (pump start, zero valve on). The flow rate is adjusted by
changing the cassette (no. 1).
CAUTION
When changing cassettes make sure that the tubes are reconnected properly.
Flow to be adjusted
Constr. No. (see figure 16)
Nominal value (tolerance)
ml/min
sampling flow
6
200 (180 to 220)
side flow
3
10 to 27
O
2
measurement in
4 and 5
45 to 60
O
2
reference in
2
25 to 42
CO
2
absorber flow
1
180 to 220 when zeroing
NOTE: Changing any of the cassettes will have some effect on the other flow rates. After any
adjustments check the other flow rates as well.
O
2
measurement flow pressure
Gradual decrease of main flow rate due to the water trap filter clogging can be checked by
measuring pressure difference between the O
2
measurement flow and the O
2
reference flow.
Remember that the sampling line should be attached to the water trap before starting the test.
The pressure difference is automatically checked after every gas zeroing.
See ACX Service menu chapter later in this manual for further information.
3.4.3
Oxygen measurement unit adjustments
The only field service procedures for the O
2
measuring unit are the offset (zero), gain, and frequency
adjustments. In case of any other trouble, the measuring unit should be replaced and the faulty
one sent to Datex-Ohmeda for repair.
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......