S/5 Airway modules
Document No. 800 1005-1
31
3
SERVICE PROCEDURES
3.1
General service information
Field service of the airway modules is limited to replacing faulty circuit boards or mechanical parts.
The circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full
type designation, and a detailed fault description.
CAUTION
Only trained personnel with appropriate equipment should perform the tests and
repairs outlined in this section. Unauthorized service may void warranty of the unit.
NOTE: After any component replacement see chapter
CAUTION
The ACX-200 photometer and its components are repaired/calibrated at the
factory. Attempts to repair/calibrate the unit elsewhere will adversely affect
operation of the unit. Datex-Ohmeda supplies spare ACX-200 photometers. The
information provided for the ACX-200 is for reference only.
CAUTION
Due to the complicated and sensitive mechanical construction any service inside
the O
2
measuring unit should not be attempted.
CAUTION
The ACX-200 Measuring board can be repaired and calibrated only at the factory.
CAUTION
The PVX-100 measuring unit can be repaired only at the factory.
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......