SPECIFICATIONS
1.1
General specifications
Module size, W
×
D
×
H
37
×
180
×
112 mm / 1.5
×
7.1
×
4.4 in
Module weight
0.37 kg /0.8 lbs
Power consumption
3.3 W
1.2
Technical specifications
1.2.1
NMT
Stimulation modes
Train of four (TOF)
Double burst (3.3) (DBS)
Single twitch (ST)
50 Hz t post tetanic count (PTC)
Measurement intervals for TOF and DBS
Manual; 10 s, 12 s, 15 s, 20 s, 1 min, 5 min, 15 min
Measurement intervals for ST Manual; 1 s, 10 s, 20 s
1.2.2
Stimulator
Stimulus pulse
Square wave, constant current
Pulse width
100, 200 or 300
µ
s
Stimulus current range
10...70 mA with 5 mA steps
Stimulus current accuracy
10 % or
±
3 mA whichever is the greater
Max. load
3 k
Ω
Max. voltage
300 V
1.2.3
Regional block mode
Stimulation modes
Single twitch
Intervals
1 s, 2 s, 3 s
Stimulus pulse
Square wave, constant current
Pulse width
40
µ
s
Stimulus current range
0...5.0 mA with 0.1 mA steps
Stimulus current accuracy
20 % or 0.3 mA whichever is the greater