TROUBLESHOOTING
4.1
Troubleshooting chart
Trouble
Cause
Treatment
Check the stimulus electrodes.
EMG electrode off.
Loose electrodes or loose stimulus
clip.
Change or attach the electrodes or
clip.
Supramax. not found.
Loose electrodes or loose stimulus
clip.
Stimulus electrodes attached to
wrong place.
Patient is relaxated.
Change or attach the electrodes or
clip.
Change the place of the stimulus
electrode.
Response too weak.
Loose stimulus electrodes.
Measuring electrodes attached to
wrong place.
Patient is relaxated.
Change or attach the electrodes.
Change the place of the meas.
electrodes.
Ref. not stable.
Patient is relaxated.
Movement artifact.