6.2.5 ECG measurement
17.
Enter the service menu:
Menu -
Monitor Setup - Install/Service
(Password 16-4-34)
-
Service View
(Password 26-23-8)
Take down the information regarding monitor software and unit id by selecting
Scroll Vers
and turning the ComWheel or choose
Record Vers
if the recorder is installed.
"
18. Enter
the
ESTP : ECG
service menu:
Menu -
Monitor Setup - Install/Service
(Password 16-4-34)
-
Service View
(Password 26-23-8)
- Modules - ESTP : ECG
Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not
increasing faster than by 50 per second. Check also that the ECG/RESP board memories
have passed the internal memory test, i.e. the “RAM”, “ROM” and “EEPROM” state all OK.
19.
Check that the power frequency value has been set according to the current mains power
frequency. Change the setting by selecting
Power Freq
, if necessary.
20.
Connect a 5-lead ECG cable to the module. Check that the “Cable type” shows 5 lead. If it
shows 3 lead, make sure the used 5-lead ECG cable contains the necessary wiring for cable
recognition (pins 0, 8 and 9 connected together).
21.
Check that all the electrodes show OFF in the service menu and the message “Leads Off” is
shown on the screen.
Connect all the leads together, for example to a suitable screwdriver. Check that all the
electrodes show ON and the message “Asystole” appears.
Disconnect one of the leads and check that the corresponding electrode in the service
menu shows OFF within 10 seconds from the disconnection, then reconnect the lead. Check
the rest of the leads using the same method.
NOTE: When the ground lead (black) is disconnected all the electrodes should show OFF.
NOTE: The asystole and different leads off -messages are shown using certain priority, so
even if only
one of the leads is disconnected, the lead related leads off -message may not
appear on the screen.
"
22.
Connect a 3-lead cable to the monitor.
Connect the leads together (e.g. to a screwdriver).
Select one lead (I, II or III) for display.
Detach a lead from the screwdriver, and see that the monitor responds as described in the
table below. Connect the lead again and check that “Leads off” message disappears.
Содержание F-LM1
Страница 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Страница 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Страница 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Страница 78: ......
Страница 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Страница 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Страница 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Страница 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Страница 94: ......
Страница 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Страница 111: ...Service menu Document No M1051339 15 4 SERVICE MENU There is no service menu available for the CPU board ...
Страница 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Страница 114: ......
Страница 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Страница 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Страница 136: ......
Страница 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Страница 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Страница 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Страница 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Страница 206: ......
Страница 208: ......
Страница 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 224: ......
Страница 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 232: ......
Страница 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Страница 244: ......
Страница 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Страница 253: ...Service menu Document No M1051340 7 6 SERVICE MENU There is no service menu available for the power adapter ...
Страница 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Страница 256: ......
Страница 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Страница 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Страница 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Страница 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Страница 286: ......
Страница 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Страница 304: ......
Страница 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Страница 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Страница 325: ...Spare parts Document No M1051358 19 ...
Страница 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Страница 330: ......
Страница 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Страница 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Страница 342: ......
Страница 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Страница 350: ...Datex Ohmeda S 5 Light Monitor Data card Figure 7 Only Data card no menu card allowed Document No 896552 2 6 ...
Страница 352: ......
Страница 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Страница 360: ......
Страница 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Страница 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Страница 368: ......