Architecture
Document No. M1051339
17
3.4 NIBP unit
The NIBP unit consists of the NIBP board, an air pump, and pneumatic valves. The NIBP board
functions can be summarized as follows:
•
Control the function of the air pump and the pneumatic valves.
•
Measure the pressure variation of the NIBP cuff.
The pressure variation of the NIBP cuff is analyzed on the NIBP board and transmitted to the CPU
board over a serial communication channel.
3.5 ECG unit
The ESTP board functions can be summarized as follows:
•
Measure ECG
•
Generate defibrillation synchronization signal
The ECG signal is monitored and analyzed on the ECG board and transmitted to the CPU board over
a serial communication channel.
The ECG board is electrically patient isolated. The operating voltage 10...16 V is generated from the
+ 12 V received from the power supply board. The ECG board is reset by the CPU board.
The ECG board generates a defibrillation synchronization signal. The signal is transmitted via the
CPU and power supply board to the serial I/O connector on the monitor rear panel. The
defibrillation synchronization signal is available also on the patient connector on the patient
connector panel.
3.6 STP unit
The STP board functions can be summarized as follows:
•
Measure SpO
2
, temperature and invasive blood pressure.
The signals from the parameters being monitored are analyzed on the STP board and transmitted to
the CPU board over a serial communication channel.
The STP board is patient isolated. The operating voltage is generated from the +12 V voltage
received from the power supply board. The STP board is reset by the CPU board.
3.7 CO
2
unit
The CO
2
unit consists of the CO
2
board, an isolated power supply, and a pump. The CO
2
board
functions can be summarized as follows:
•
Process the CO
2
waveform generated by the mainstream CO
2
sensor.
•
Generate end tidal and inspiratory CO
2
values, and breaths per minute data.
•
Control the Mainstream CO
2
sensor motor and heater.
The waveform data is processed on the CO
2
board and transmitted to the CPU board over a serial
communication channel.
Содержание F-LM1
Страница 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Страница 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Страница 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Страница 78: ......
Страница 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Страница 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Страница 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Страница 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Страница 94: ......
Страница 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Страница 111: ...Service menu Document No M1051339 15 4 SERVICE MENU There is no service menu available for the CPU board ...
Страница 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Страница 114: ......
Страница 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Страница 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Страница 136: ......
Страница 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Страница 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Страница 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Страница 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Страница 206: ......
Страница 208: ......
Страница 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 224: ......
Страница 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 232: ......
Страница 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Страница 244: ......
Страница 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Страница 253: ...Service menu Document No M1051340 7 6 SERVICE MENU There is no service menu available for the power adapter ...
Страница 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Страница 256: ......
Страница 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Страница 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Страница 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Страница 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Страница 286: ......
Страница 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Страница 304: ......
Страница 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Страница 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Страница 325: ...Spare parts Document No M1051358 19 ...
Страница 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Страница 330: ......
Страница 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Страница 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Страница 342: ......
Страница 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Страница 350: ...Datex Ohmeda S 5 Light Monitor Data card Figure 7 Only Data card no menu card allowed Document No 896552 2 6 ...
Страница 352: ......
Страница 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Страница 360: ......
Страница 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Страница 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Страница 368: ......