Service menu
5.1 NIBP service menu
Service data
Pressure
shows measured pressure multiplied by 10.
Zero
shows pressure at auto zeroing multiplied by 10 and changes b20 and -20 mmHg.
Absolute pressure is the sum of
Pressure
and
Zero
.
Protect handle
indicates hardware protection for EEPROM memory. It should be ON all the time in
normal operation. If it is OFF, data cannot be read from or written to EEPROM, only the calibration
protection can be set or reset by software. It can be turned to OFF by turning the toggle switch to the
right at the bottom of the module, which also enables 'Protection ON/OFF' menu selection in the
calibration menu.
Calibr. prot.
shows software calibration protection and should be OFF to enable calibration.
+12 V power
indicates the condition of the supply v12 VD for the pump and valves. It
exists (ON) or not (OFF) depending on service menu function.
AD0
to
AD7
show the values of each eight channels of A/D converter.
Timeouts
is a cumulative number that indicates how many times the unit has not responded to the
CPU's inquiry.
Bad checksums
is a cumulative number that indicates how many times
communication from the unit to CPU has broken down.
Bad c-s by mod
is a cumulative number that indicates how many communication errors the unit
has detected.
The Light Monitor starts counting these items at power up and resets to zero at power off. The
nonzero values do not indicate a failure, but the continuous counting (more than 50 per second)
indicates serial communication failure. Also other units can cause communication errors that
cause these numbers rise.
RAM
indicates the state of the external RAM memory.
ROM
indicates whether the checksum in the EPROM is in accordance with the one the software has
calculated.
EEPROM
indicates if the values stored in the permanent memory are valid.
The state is either
OK, Fail
or
?
(unit not in place or a communication error).
Document No. 8000060-1
51
Содержание F-LM1
Страница 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Страница 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Страница 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Страница 78: ......
Страница 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Страница 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Страница 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Страница 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Страница 94: ......
Страница 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Страница 111: ...Service menu Document No M1051339 15 4 SERVICE MENU There is no service menu available for the CPU board ...
Страница 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Страница 114: ......
Страница 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Страница 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Страница 136: ......
Страница 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Страница 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Страница 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Страница 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Страница 206: ......
Страница 208: ......
Страница 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 224: ......
Страница 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 232: ......
Страница 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Страница 244: ......
Страница 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Страница 253: ...Service menu Document No M1051340 7 6 SERVICE MENU There is no service menu available for the power adapter ...
Страница 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Страница 256: ......
Страница 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Страница 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Страница 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Страница 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Страница 286: ......
Страница 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Страница 304: ......
Страница 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Страница 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Страница 325: ...Spare parts Document No M1051358 19 ...
Страница 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Страница 330: ......
Страница 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Страница 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Страница 342: ......
Страница 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Страница 350: ...Datex Ohmeda S 5 Light Monitor Data card Figure 7 Only Data card no menu card allowed Document No 896552 2 6 ...
Страница 352: ......
Страница 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Страница 360: ......
Страница 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Страница 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Страница 368: ......