![Datex-Ohmeda AS/3 Compact Monitor Скачать руководство пользователя страница 82](http://html.mh-extra.com/html/datex-ohmeda/as-3-compact-monitor/as-3-compact-monitor_technical-reference-manual_513828082.webp)
Datex-Ohmeda AS/3 CM and CS/3 CMC
74
Document No. 896 623
6
GENERAL TROUBLESHOOTING
Monitor not functioning
STBY -LED
lit ?
Start-up texts
appear on
screen?
Clock etc
appears on
the screen?
Plug in a module
Module data
appear on
the screen?
Another
module works
in same place?
Yes
Remove all modules
and turn power on
Yes
Yes
Yes
Yes
Faulty module. Go to Module
troubleshooting in part II.
Faulty Central Unit. Go to
troubleshooting in part II.
Faulty Central Unit. Go to
troubleshooting in part II.
Pick up next module
Connect and check
OK?
No
No
No
No
No
No
No
Keyboard trouble. See part II
for troubleshooting.
The power supply unit
was shut down by instant over
voltage in the mains.
Yes
Power cord
connected?
On/STBY switch ’ON’
Display connected?
Keyboard connected?
OK?
Yes
Disconnect and reconnect the power cord
No
Connect power cord
OK?
No
No
Yes
Figure 15
Monitor General Troubleshooting Flowchart