![Datex-Ohmeda AS/3 Compact Monitor Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/datex-ohmeda/as-3-compact-monitor/as-3-compact-monitor_technical-reference-manual_513828066.webp)
Datex-Ohmeda AS/3 CM and CS/3 CMC
58
Document No. 896 623
5.3
Functional Inspection
Start up
9.
Connect the mains power cord to the Frame.
Check that the stand-by LED is lit .
?
10.
Check that the fans start running.
?
11.
Turn the monitor on. Check that the monitor starts up properly, i.e.
−
the alarm LED’s turn on and off
−
the start-up sound is heard from the loudspeaker
−
the normal monitoring screen appears
−
no error messages appear onto the screen
−
If the monitor contains a recorder, two lines of start-up information should be
recorded.
?
Monitor setup
12.
Configure the monitor screen so that all required parameter information is displayed:
Monitor Setup
- Screen 1 Setup - Waveform Fields
- Digit Fields
Check that all of the connected modules are recognized, i.e. the required parameter
information is displayed on the screen.
NOTE: The InvBP waveform is not displayed if the InvBP cable is not connected
If some parameter information is missing, check that the interfacing settings are correct:
Monitor Setup -
Install/Service
(Password 16-4-34)
Installation - Interfacing.
?
13.
Preset the measurement settings for the connected parameters, for examples:
Monitor Setup
- Screen 1 Setup - Digit Fields -Lower Field 4 - PgCO
2
Others
- EEG - EEG Setup - Numeric 1 - MF
- Numeric 2 - Ampl.
- Montage - EEG Channels - 4
- Montage type - Bip
-
EP - Cycle - Cont.
- EP Size - 1
- AEP Setup - AEP Channels - 2