General Service Guide
53
Document No. 896 623
SpO
2
>40, SpO
2
>60, SpO
2
>80 (beat-to-beat, display value, 10 s average): The original scale 40-
100% (SpO
2
>40), 60-100% (SpO
2
>60) or 80-100% (SpO
2
>80) is scaled between -5 and +5
volts.
CO
2
: The original scale 0%...10% is scaled between 0 and +5 volts. Values greater than 10% are
set to 10%. (Airway gas special indications are applied, see also chapter Special Indications).
AA (Anesthesia Agent): The original scale 0%...10% is scaled between 0 and +5 volts. Values
greater than 10% are set to 10%. (Airway gas special indications are applied, see also chapter
Special Indications).
O
2
: The original scale 0%...100% is scaled between 0 and +5 volts. (Airway gas special indications
are applied, see also chapter Special Indications).
N
2
O: The original scale 0%...100% is scaled between 0 and +5 volts. (Airway gas special
indications are applied, see also chapter Special Indications).
Paw (Airway Pressure): The original scale -20 cmH2O...+80 cmH2O is scaled between -5 and +5
volts. (Airway gas sensor failure is applied, see also chapter Special Indications).
Flow: The original scale -100 l/min...+100 l/min is scaled between -5 and +5 volts. (Airway gas
sensor failure is applied, see also chapter Special Indications).
Volume: The original scale -2.5 litters...+2.5 litters is scaled between -5 and +5 volts. (Airway gas
sensor failure is applied, see also chapter Special Indications).
Resp: The original scale -5000 mohms is scaled between -5 and +5 volts.
RR: RR display 0 to 150 breaths per minute are scaled between 0 and +1.5 volts.
T1, T2, T3, T4 (Temperature): The original scale 0 degrees...50 degrees Celcius is scaled between 0
and +5 volts.
TEST SIGNALS -5 V, 0 V, +5 V: Steady signals with one of the listed values.
TEST 1: Test signal of a triangle shape with base width of 1 second, min value of 0 volt and max
value of 5 volts.
TEST 2: Test signal of a triangle shape with base width of 4 seconds, min value of -5 volts and max
value of +5 volts.
4.8.3
Setting the Analog Output Signals
Set the analog output signals according to the following procedure.
1.
Press the
Monitor Setup
key .
2.
Select
Install/Service
and enter the password (16-4-34) to get into the corresponding
menu.
3.
Select
Analog Outputs
to get into the corresponding menu and combine a required channel
with an output signal.
4.
Press the
Normal Screen
key.