S/5 Anesthesia Record Keeping Keyboard and Barcode Reader
Document No. 8001019-3
11
Press the keys on the lower part, all except the keys
Modify
and
. Check that each key
generates a sound from the loudspeaker, or at least the ‘Message count’ value increases in
the service menu.
Press the keys
Modify
and
and check that the corresponding menus open onto the
screen.
?
9.
Perform electrical safety check and leakage current test.
?
10.
Check that the Keyboard for the deioRecorder for Anesthesia functions normally after the
performed electrical safety check.
?
11.
Clean the Keyboard for the deioRecorder for Anesthesia with suitable detergent.
?
•
Fill in all necessary documents.
3.3
Disassembly and reassembly
The Keyboard for the deioRecorder for Anesthesia, K-ARKB is disassembled in the following way.
See the exploded view of the keyboard.
1.
Disconnect the ARK keyboard - monitor cable (ARK keyboard - LCD display cable).
2.
Remove four screws from the bottom of the keyboard, and detach the cover plate from the
bottom plate.
Содержание ARK N-SCAN
Страница 2: ......
Страница 4: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ii SERVICE CHECK FORM A 1 ...
Страница 26: ...Datex Ohmeda S 5 monitors Document No 8001019 3 22 ...
Страница 27: ...APPENDIX A Service check form K ARKB Document No 8001019 3 23 APPENDIX A ...
Страница 28: ...Datex Ohmeda S 5 monitors Document No 8001019 3 24 ...
Страница 30: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ...