Datex-Ohmeda ADU
C-54
Document No. 8502855
*) Check also Ventilator failure alarm troubleshooting guide for reason for ventilator failure alarm.
Beware of the O
2
supply lost alarm and the ventilator communication alarm will go off as a
consequent of a ventilator failure alarm.
Ventilator failure alarm troubleshooting guide
The suggested actions below should be seen as a starting point when a intermittent Ventilator failure
alarm has occurred, this means that an sensor error for example can be due to a broken CPU board or
DC/DC board.
The reason for a ventilator failure (Safe State) is stored in the Ventilator EEPROM. In position 15 an ID
is stored and in position 16 to 19 some extra information is stored. All values are stored in
hexadecimal format.
Beware of as long the Ventilator failure alarm is present, can not the correct data be presented from
the EEPROM on the A-EV1 Diagnostic page. This because the communication between A-ELEC and A-
EV1 not are working correctly (this is also the reason to that the O
2
supply lost alarm and Ventilator
communication error alarm comes as sub alarms to the Ventilator failure alarm).
To check the EEPROM values on a failing ADU with a Ventilator failure alarm must the ADU be
restarted and no Ventilator failure alarm be present.
If the alarm is present should the fault be traced by entering the A-EV1 diagnostic page and
performing the trouble shooting from the messages found under STATUS in the diagnostic page.
Ventilator
*
failure
HW watchdog err
HW watchdog circuit
does not reset system
during internal system
check
Replace CPU board
x
Ventilator
*
failure
ROM check sum
Program memory is faulty Replace program
or
replace CPU board
x
Ventilator
*
failure
RAM W/R error
Faulty components on
CPU board
Replace CPU board
x
Ventilator
*
failure
Hw reset
Unexpected reset of CPU
Replace CPU board
Ventilator
*
failure
Flow zero-error
Failed test of flow zeroing
valve
Replace flow sensor or
zeroing valve
Ventilator
*
failure
Backup CPU Error
Failed test of backup CPU Replace CPU board
Ventilator
*
failure
pilot press err
During the A-EV1 Self-
test was the pilot
pressure below
120cmH
2
O.
Calibrate pilot pressure
regulator.
Check pipeline gas
supply pressure.
Replace pilot pressure
regulator
x
Message
in
Normal
Screen
Message in
diagnostic
page
Error condition /
possible cause
Recommended
action
Err
log
Содержание Anesthesia Delivery Unit S/5
Страница 2: ......
Страница 14: ......
Страница 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Страница 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Страница 76: ...Datex Ohmeda ADU 5 18 Document No 8502852 Figure 24 Transformer unit Figure 25 Transformer unit 2 1 4 6 ...
Страница 80: ...Datex Ohmeda ADU 5 22 Document No 8502852 Figure 30 Compact Circuit A CC1 exploded view 5 1 2 3 4 ...
Страница 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Страница 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Страница 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Страница 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Страница 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Страница 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Страница 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Страница 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Страница 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Страница 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Страница 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Страница 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Страница 148: ......
Страница 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Страница 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Страница 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Страница 211: ...Spare parts A AUF Document No 8502853 A 59 Figure 53 Exploded view of the gas inlet block and the gas labels 4 1 2 3 ...
Страница 229: ...Technical description A FGC1 Document No 8502854 B 3 Figure 3 Technical principles of the A FGC1 ...
Страница 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Страница 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Страница 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Страница 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Страница 307: ...Spare parts A FGC1 Document No 8502854 B 81 Figure 60 Fresh Gas Control Unit A FGC1 1 2 4 3 6 5 7 8 10 11 12 ...
Страница 311: ...Spare parts A FGC1 Document No 8502854 B 85 Figure 64 Parts inside the A FGC1 1 2 3 4 5 6 7 8 10 ii iii i iv 9 ...
Страница 319: ...Spare parts A FGC1 Document No 8502854 B 93 Figure 71 Gas connection unit 1 2 3 4 5 6 7 8 9 10 11 ...
Страница 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Страница 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Страница 362: ...Datex Ohmeda ADU C 36 Document No 8502855 The ventilator calibrating tree Figure 27 The ventilator calibrating tree ...
Страница 400: ...Datex Ohmeda ADU C 74 Document No 8502855 Ventilator diagnostic page layout Figure 42 Ventilator diagnostic page ...
Страница 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Страница 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Страница 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Страница 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Страница 477: ...Troubleshooting A ELEC Document No 8502856 D 47 Trouble shooting PCMCIA Card Figure 57 Troubleshooting PCMCIA Card ...
Страница 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Страница 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Страница 486: ...Datex Ohmeda ADU D 56 Document No 8502856 Figure 63 Transformer unit Figure 64 Transformer unit 2 1 4 ...
Страница 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...