1 Introduction
1006-0453-000 05/04
1-3
1.2 Standard service procedures
1.2.1 Operation
manuals
You must have, and be familiar with, the Operation manuals for this product.
Study the Aestiva Operation manuals if you need further information about the
operation of the system.
1.2.2 Service
manuals
You must determine where a problem is located before you can determine
which service manual to use:
•
Use this manual for 7900 Ventilator related problems.
•
Use the Anesthesia Machine service manual (1006-0452-000)
for all other components of the Aestiva Anesthesia Machine.
1.2.3 Ventilator
tests
Service calibration functions let Datex-Ohmeda trained users and Datex-
Ohmeda service personnel perform ventilator setup functions, tests,
calibration and measurements from the front panel display.
Normal operational tests, calibration, and troubleshooting can be performed
on your Aestiva 7900 Ventilator without removing components from the
system. Repair may require removing the ventilator components from the
anesthesia machine.
w
w
w
w
WARNING
Section 4,
“Service Mode Tests and Calibration”
must be performed
whenever you access any internal component of the Ventilator to verify
that all critical parts of the Ventilator are still operational and within
specification.
w
w
w
w
WARNING
After the Ventilator has been serviced, you must perform
“Post -Service
Checkout”
to verify the entire Anesthesia System is properly functioning
before the system can be returned to clinical use.
w
w
w
w
WARNING
Do not perform testing or maintenance on this instrument while it is
being used to ventilate a patient, possible injury may result.
Содержание Aestiva 7900 SmartVent
Страница 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Страница 44: ...Notes 2 28 05 04 1006 0453 000 ...
Страница 81: ...4a Tests and Calibration 4 X Software 1006 0453 000 05 04 4a 35 Figure 4 38 Service Calibration full menu ...
Страница 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Страница 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Страница 150: ...Notes 5 34 05 04 1006 0453 000 ...
Страница 158: ...Notes 6 8 05 04 1006 0453 000 ...
Страница 182: ...Notes 7 24 05 04 1006 0453 000 ...
Страница 189: ...8 Illustrated Parts 1006 0453 000 05 04 8 7 U14 U23 4 12 13 14 15 16 15 18 21 20 23 24 1 1a 22 17 19 ...
Страница 202: ...Notes 8 20 05 04 1006 0453 000 ...
Страница 203: ......