When to change the
absorbent
A gradual color change of the absorbent in the canister indicates
absorption of carbon dioxide. The color change of the absorbent is
only a rough indicator. Use carbon dioxide monitoring to determine
when to change the canister.
Discard the absorbent when it has changed color. If left standing for
several hours, absorbent may regain its original color giving a
misleading indication of activity.
Important
Read the canister instructions completely before using the product.
Removing a canister
1. Hold the canister by the handle and push on the release latch to
unlock the canister.
2. Remove the canister by tilting it downward and off the two
support pins.
AB.
74p
05
8