Aespire View
7-6
M1132382
100% O2 cell
calibration
Complete a 21% O2 cell calibration before performing a 100% O2 cell
calibration.
1. Make sure that the O2 cell is in the circuit.
2. After performing a 21% calibration, select
100% O2
.
3. Push the O2 flush button for 5 seconds and set the O2 flow to
5 l/min and set other gases to minimal flow.
4. Select
Start Calibration
.
5. ‘Complete’ shows on the screen upon successful calibration.
6. If ‘Failure’ shows on the screen, repeat the 100% O2 cell
calibration.
•
If the calibration fails after another attempt, decrease the
airway pressure and try the 100% O2 cell calibration again.
•
If calibration fails after repeated attempts, perform a 21% O2
cell calibration. Then try a 100% O2 cell calibration again.
•
If the 100% O2 cell calibration does not pass, replace the O2
cell.
7. Perform the
in the
section before using the system.
Zeroing flow sensor
WARNING
Do not perform calibration while system is connected to a
patient.
Note
The system automatically corrects for zero offset when the flow
sensor connectors are unplugged and the system power is on.
1. Set the Bag/Vent switch to Bag.
2. Remove the flow sensor module.
•
Pull the latch to unlock the flow sensor module from the
breathing system.
•
Pull the flow sensor module out of the breathing system.
3. ‘No insp flow sensor’ and ‘No exp flow sensor’ show on the
display when the zeroing is complete.
4. Reinstall the flow sensor module.
5. Perform the
in the
section before using the system.”
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...