6 Alarms and Troubleshooting
M1132382
6-13
Electrical problems
WARNING
If a circuit breaker opens frequently, do not use the
system. Have a Datex-Ohmeda trained service
representative repair the system.
System
Problem
Solution
Mains indicator is not on. The electrical power cable is not
connected.
Connect the power cable.
The inlet circuit breaker (switch)
is off.
Turn the circuit breaker on.
The power cable is damaged.
Replace the power cable.
The electrical socket the power
cable connects to had no power.
Use a different electrical
socket.
An internal fuse is open.
Have a Datex-Ohmeda trained
service representative repair
the system.
One electrical outlet does
not have power.
The outlet circuit breaker is off.
Turn the circuit breaker on.
A circuit breaker opens
frequently.
Equipment connected to the
outlet uses more current than
the circuit breaker rating.
Use a different power supply
for some of the equipment.
The equipment connected to the
outlet has a short.
Have a Datex-Ohmeda trained
service representative repair
the system.
Tec 6 series vaporizer
has no power.
Not plugged into outlet.
Connect power cable.
The outlet circuit breaker is off
Turn the circuit breaker on.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...