Aespire View
6-4
M1132382
List of alarms
If corrective action does not resolve the alarm, contact a
Datex-Ohmeda trained service representative.
Message
Priority
Cause
Action
+15V Analog out-
of-range
Minimum shutdown
(High)
Ventilator malfunction.
Ventilate manually. Monitoring is not reliable.
Contact a Datex-Ohmeda trained service
representative.
+6V Analog out-of-
range
Minimum shutdown
(High)
Ventilator malfunction.
Ventilate manually. Monitoring is not reliable.
Contact a Datex-Ohmeda trained service
representative.
-6V Analog out-of-
range
Minimum shutdown
(High)
Ventilator malfunction.
Ventilate manually. Monitoring is not reliable.
Contact a Datex-Ohmeda trained service
representative.
5V Ref out-of-range Minimum shutdown
(High)
Ventilator malfunction.
Ventilate manually. Monitoring is not reliable.
Contact a Datex-Ohmeda trained service
representative.
1.225 V Ref out-of-
range
Minimum shutdown
(High)
Ventilator malfunction.
Ventilate manually. Monitoring is not reliable.
Contact a Datex-Ohmeda trained service
representative.
12 hour test
recommended
Informational
System is in use for more than 12
hours without a power-up self test.
At end of the case, move the System switch
from On to Standby to On.
A/D converter
failure
Minimum shutdown
(High)
Ventilator malfunction.
Ventilate manually. Monitoring is not reliable.
Contact a Datex-Ohmeda trained service
representative.
Absorber panel
open
Medium
The breathing system is not fully
latched.
Fully latch the breathing system.
ACGO on
Medium
(Informational after
acknowledge)
The outlet selection switch is set to
the auxiliary common gas outlet.
Connect the patient circuit to the auxiliary outlet.
For mechanical ventilation or ventilation with
monitoring, select the circle system setting.
Apnea > 120 s
High
No mechanical breaths or
spontaneous breaths greater than 5
ml in last 120 seconds.
Check the patient. Bag as needed. Check for
disconnects. If the patient is on a heart lung
machine, select Cardiac Bypass from the Main
menu.
Apnea
Medium
No mechanical breaths or
spontaneous breaths greater than 5
ml in last 30 seconds.
Check the patient. Bag as needed. Check for
disconnects. If the patient is on a heart lung
machine, select Cardiac Bypass on the Main
Menu.
Apnea alarm off
Informational
The Cardiac Bypass option is set to
On in the Main Menu.
Set Cardiac Bypass to Off in the Main Menu.
Apnea alarm
standby
Informational
Normal condition after End Case,
power-up, or ACGO change from On
to Off.
Monitoring resumes after first breath
(mechanical) or two breaths within 30 seconds
(non-mechanical).
Backup mode
active
Informational
No spontaneous breaths in set period
of time and 30 seconds have elapsed
since starting PSVPro mode.
Select a new ventilation mode or switch to
manual ventilation.
Battery charger fail
Informational
The current in the battery charging
circuit is too high.
System is operational, but may fail on battery if
mains power is lost. Contact a Datex-Ohmeda
trained service representative.
Battery charging
Informational
Battery is not fully charged.
Leave the system plugged in to charge the
battery.
Battery circuit
failure
Informational
Battery measures less than 7 V or
higher than 16.5 V.
System is operational, but may fail on battery if
mains power is lost. Contact a Datex-Ohmeda
trained service representative.
Battery current high Informational
Battery current greater than 6 amps
for 10 seconds.
System is operational, but may fail on battery if
mains power is lost. Contact a Datex-Ohmeda
trained service representative.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...