7100 Anesthesia Ventilator
4-22
02/03 1006-0836-000
4.9.8
Service Calibrations
Required
The
Service Calibrations Required
W
menu displays which setting or
calibration must be performed when the “Service Calibration
w
” alarm
appears in normal operation. After the setting or calibration is properly
completed, the text for that setting or calibration will be removed.
Remarks
The normal operation “Service Calibration
w
” alarm message is only removed
when all the required settings or calibrations are completed.
The
Set
“
Service Calibration
W
” menu item is used by the factory to activate
the “Service Calibration
w
” alarm and require that all settings and calibrations
be performed when the machine is set up for operation at its permanent
location.
You can reset the “Service Calibration
w
” alarm in the field by selecting
“Yes” when the following warning appears after selecting the
“Set Service Calibration
W
” menu item.