7100 Anesthesia Ventilator
4-18
02/03 1006-0836-000
4.9.6
Inspiratory Valve
Calibration
The
Inspiratory Valve Calibration
should be performed:
•
when the machine is first put into service.
•
at prescribed, planned maintenance intervals.
•
after the pneumatic engine has been serviced.
Calibration setup
1. Remove the bellows and pressure relief valve (pop-off) from the bellows
assembly.
•
Remove the bellow housing.
•
Remove the bellows and bellows mounting rim as an assembly.
•
Remove the pressure relief valve.
•
Replace the bellows housing.
•
Note:
If you have just completed the PEEP Valve Calibration, remove
these parts from the bellows assembly and then reassemble the
breathing system.
2. Attach a patient circuit tube to the Calibrated Flow Orifice test tool.
3. Connect the open end of the patient circuit tube to the inspiratory flow
patient connection.
4. Set the Bag/Vent switch to Vent.
5. Establish drive gas for the ventilator.
6. Adjust fresh gas flow so the actual Airway Pressure reads 1.05 cm H
2
O.