7100 Anesthesia Ventilator
7-8
02/03 1006-0836-000
7.2.4 Front enclosure with
control board removed
To replace the speaker
1. Remove the two screws (
A
) that hold the speaker assembly (
B
) to the front
enclosure.
2. Reassemble in reverse order.
To remove the front enclosure
1. Use a 7-mm nut driver to remove the twelve nuts (
C
) around the periphery
that hold the front enclosure (
D
) and the front panel assembly (
E
) together.
2. Carefully lift the front enclosure from the front panel assembly. Guide the
cables through the slotted openings at the left edge of the front enclosure.
3. Reassemble in reverse order.
A
D
B
C
E