7 Repair Procedures
1006-0836-000 02/03
7-7
7.2.3 Battery and
power supply
w
w
w
w
Caution
The circuit boards are electrostatic sensitive. Use an anti-static
workstation and wear a wrist grounding strap when handling a circuit
board.
To replace the Battery
1. Disconnect the battery cable (
A
) from the control board.
2. Note the orientation of the wires that connect to the battery and how they
route through the retaining clamp.
3. Remove the battery’s mounting strap (
B
)
4. Replace the battery in the same orientation noted above (wires facing
inward.
To replace the Power Supply
1. Disconnect the battery cable (
A
) from the control board.
2. Disconnect the power supply’s output connector (
C
).
3. Disconnect the power supply’s input connector and ground wire (
D
).
4. Loosen each mounting screw (four corners) until it begins to ratchet.
5. Lift out the power supply.
6. Reassemble in reverse order.
Ground Terminal
Note the sequencing of ground wires and hardware when replacing any of the
harnesses.
Chassis
L
PI
PS
CB
L
L
N
N
PI - Ground lug from Power Inlet
PS - Ground lug from Power Supply
CB - Ground lug from Control Board
L - Lockwasher
N - Keps nut
A
B
C
D
Note:
After replacement, the
battery must be charged until
Battery Status is indicated as
“Battery Charged.”
Refer to Section 4.10.3.