7 Repair Procedures
1006-0836-000 02/03
7-3
7.1.1 After replacing
the Control Board
or the Control Module
1. Load the appropriate software revision using a software memory stick that
matches the software revision of the replaced control board (control
module).
2. Enable/Disable the appropriate ventilator settings in the
Service Mode,
System Configuration
menu, Section 4.8, to match the replaced control
board.
3. Install new control board configuration label on back of control module
(with Control Board BID number and Feature KEY code). The replacement
control module has the label already attached.
143AB99
143AB99
Replacement Control Board