7100 Anesthesia Ventilator
5-20
02/03 1006-0836-000
5.4.6 No ventilation
troubleshooting
* Refer to Section 7.2,
“Breathing System Leak Tests,”
in the respective anesthesia machine Technical Reference Manual.
Yes
No
Yes
Start
Zero Flow and
airway pressure
transducers
Verify pass?
High flow Gas
exhausts from
drive gas port?
No
1. Remove bellows assembly
2. Enter Valves - Test Tool
3. Turn on the flow valve to 15 L/min
4. Check regulator calibration
Yes
Correct Hospital
Supply Problem
Perform airway pressure
transducer calibration or
proceed to Inaccurate
Ventilation Troubleshooting
Section 5.4.3
Yes
No
1. Turn off the flow valve
2. Turn on the PEEP Safety Valve
3. Set the PEEP valve to 100 cm H2O
4. Using hand, occlude bleed resistor
on bottom of vent engine
Gas exhausts from
PEEP control port?
Regulator
pressure OK
at 25 psi?
Set Flow valve
to 70 L/min
Check hospital
pressure supply
Hospital supply
pressure OK at
35-100 psi?
Replace Regulator
and perform all
valve calibrations
Yes
No
Inspiratory Flow Valve Problem
1. Check electrical connections
2. Attempt inspiratory flow valve
calibration
3. Replace inspiratory valve if
necessary and recalibrate
Setup for PEEP Valve
Calibration but use the
Valve test tool menu:
With Peep valve set to
100 cm H
2
O
Yes
Pressure
90-110 cm H
2
O?
1. Examine Exhalation
Valve and exhalation
valve diaphragm and
reassemble breathing
circuit
2. Reduce PEEP setting
to 70 cm H
2
O
3. Place a plug in the
drive port of the
Bellows Assy. The
bellows can be left off
4. Occlude inspiratory
port and press flush
button
Yes
Does Circuit
Pressurize?
Does Bellows
Move?
No
Check all valve calibrations.
If OK, go to Breathing System
Leak Troubleshooting if bellows
does not refill
Proceed to Breathing
Circuit Leak
Troubleshooting
No
1. Close PEEP Safety Valve and
Remove PEEP Valve from the
Vent Manifold
2. Turn on PEEP Safety Valve
Does flow come from
vent engine plate?
PEEP Safety Valve Problem
1. Check electrical supply
and connections
2. Replace if necessary
No
PEEP Control Valve Problem
1. Check electrical supply
and connections
2. Attempt Recalibration
3. Replace if necessary
No
Yes
No
Check function of
the drive pressure
pop-off valve
Yes
No
Yes
Problem solved?
If not replace filter
Yes
*
*
AB.43.103