The battery inside the instrument is a special lithium battery. In order to give full play to the performance of the battery,
please use the internal battery power supply when you start using the instrument. The first use of the battery needs to be
depleted, and then charging the battery, the first charging time should be no less than 10 hours. Battery charging
temperature range is 0°C~50°C. For your safety, the charge will terminate automatically when the charging temperature is
too high. When the instrument is idle for more than two months, it should be charged in time to maintain the battery power.
Please don’t take out the battery without permission. Please do not let the battery close to the fire source or strong heat.
The battery should be removed when the instrument is stored for a long time. The storage temperature range of the
battery is -20°C~45°C.
Instrument screen cleaning and touch calibration
The display of the instrument is a 5.8 inch color LCD with touch screen. Please do not use sharp objects to click on the
screen, otherwise it may lead to damage to the LCD screen. You can clean the LCD screen using soft fabric, but do not
use organic solvent.
Instrument calibration
It is recommended that the instrument be calibrated once every two years. Please contact the supplier for specific items.
Common faults and troubleshooting
Common faults
Failure cause
Troubleshooting
Instrument can not start normally
Battery low power
Battery charging
Instrument can not charge properly
Ambient temperature is too high or low
Charge the battery in the 0
℃
~45
℃
Poor battery connect
Check battery connectors
Instrument problem
Contact supplier
Inaccuracy test results
Incorrect parameter setting
Resetting parameters
Optical fiber end contamination
Clean the end face of fiber
Connector type mismatch
Replacing matching connector
Optical port damage
Replacing fiber connector
Noise too high of test curve
Connector type mismatch
Replacing matching connector
Pulse width too small
Increasing pulse width
Incorrect attenuation
Modify attenuation
Saturation of curve front end
Pulse width too large
Reducing pulse width
Incorrect attenuation
Modify attenuation
Slow down of curve front end
Optical fiber end contamination
Clean the end face of fiber
Optical port damage
Replacing fiber connector
Connector types mismatch
Replacing matching connector
Unable to measurement
the end of fiber
Range too small
Increasing test range
Pulse width too small
Increasing pulse width
Event analysis error
Incorrect test conditions
Modify test conditions
Optical fiber length inaccuracy
Incorrect test conditions
Modify test conditions
Incorrect fiber refractive index
Modify fiber refractive index
Incorrect optical cable correction
Modify optical cable correction