List Arb
The List Arb allows output to run in sequence.
List Arb contains up to 512 current or voltage steps (Option E36150ADVU) that are entered individually in the Arb
(List) Properties window. For standard option, you can configure up to 100 voltage and current steps.
You can also populate the List Arb with values from a previously configured "standard" Arb and then edit the steps in
the Arb (List) Properties window by using Edit Points function.
Step 1– Select Arb waveform
Press Arb to access the List Arb. The List Arb is selected by default.
Step 2– Add/remove steps to/from the List
Press Add to insert a new step below the selected step. Note that the values in the new step are copied from the
previous step. Continue to add steps until your sequence is complete. Use the navigation keys to move through the
list.
Press Delete if you want to delete the selected step. Press and hold Delete if you want to remove all steps from the
list.
Keysight E36150 Series User's Guide