OCP Delay Start
Specifying an OCP Delay Start lets the OCP circuit ignore the CC status bit during the specified delay period. Once
the OCP delay time has expired and if the CC mode persists, the output will shut down.
You can specify the OCP delay start timer through:
– CC Transition: Delay timer starts at any transition of the output into CC mode. Press OCP Start CC.
– Setting Change: Delay timer starts at the end of a settings change in voltage, current, or output state. Press
OCP Start Set.
Clears and OVP and OCP Event
To clear the protection function, first remove that condition that caused the protection fault.
Press Protection Clear to clear the protection function and returns the output to its previous operating state.
From the remote interface:
To set the over-voltage protection to the maximum limit:
VOLT:PROT MAX
To enable the over-current protection:
CURR:PROT:STAT ON
To set the over-current protection delay time at 3 seconds:
CURR:PROT:DEL 3
To set the over-current protection delay timer start to CC transition:
CURR:PROT:DEL:STAR CCTR
To clear protection:
OUTP:PROT:CLE
70
Keysight E36150 Series User's Guide