View the error log
Press Utilities > Error to display the error log. If there are more than 10 errors on the display, press Next to scroll to
the next page.
Press Back or [Meter View] to return to the meter-view display.
– Errors are stored in the order they are received. The error at the end of the list is the most recent error.
– If there are more than 20 errors in the queue, the last error stored is replaced with -350,"Queue overflow". No
more errors are stored until you remove errors from the queue. If there are no errors, the instrument responds
with +0,"No error".
– Except for self-test errors, errors are cleared when exiting the Error Log menu or when cycling power.
If you suspect that there is a problem with the power supply, refer to the "Troubleshooting" section in the
Service
Guide
.
Keysight E36150 Series User's Guide
65