Installation services requirements AutoLab TF dispensers TFINREQTF / Rev 13.0 / J Fain / 1
st
Oct 10 Page 5 of 24
Space requirement for AutoLab TF systems
Please refer to below drawings which indicate the suggested space requirements in the laboratory for each type of
AutoLab TF system. There are several important items that you need to pay attention to during the positioning of
machine:
The location of main power circuit breaker is recommended to be as close to the machine as possible. A 15-
metre power cable is supplied with the machine; make sure the mains circuit breaker does not exceed this
length.
You need to keep at least 60cm width all around the machine, for easy access when servicing the machine
A 15-metre PC-to-Dispenser communication cable is supplied with the machine; make sure the location of
computer table does not exceed this range, taking into consideration the communication cable may need to be
routed up into the ceiling.
Do NOT install system near to air conditioning unit outlet, or next to a door, since this may make the scale
unstable