Datacolor CONDITIONER™ (MB2) 06/2017
page 30/126
4.6
Placement of the light cassettes (cabinet with lights)
The light cassettes are permanently connected. The cable length permits easily removal to replace the
fluorescent tubes or to change the position of the light cassette inside the cabinet.
You can insert the light cassettes in different heights onto the beads at the lateral walls of the chamber.
Insert and pull out the light cassettes only by the handles.
Do not place any charging material directly onto the light cassettes because those are heated by the fluo-
rescent tubes which would lead to exposing the charging material to undefined temperatures. The tem-
perature directly below or on the light cassettes is not equal to the temperature displayed at the tempera-
ture controller.
Place the charging material on the supplied racks below the light cassettes.
CAUTION
The light cassettes will become hot with temperature set-points >40 °C.
Danger of burning.
∅
Do NOT touch the light cassettes during operation.
Let the light cassettes cool down before changing their position.
4.7
Electrical connection
The cabinets are supplied ready for connection. They come with a fixed power connection cable of at
least 1800 mm /
70.87 in
in length.
•
Power plug: NEMA 6-20P
•
V/-10% at the indicated power frequency: 200-240 V at 50Hz, 200-240 V at 60Hz
•
Current type: 2~
•
Cabinet fuse: 16 Amp
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the cabinet’s protective conductor meets the latest
technology. The protective conductors of the socket and plug must be compatible!
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the cabinet’s type plate (left cabinet side, bottom right-hand, see chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company.
We recommend the use of a residual current circuit breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Installation category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 23.4).