18
•
Controls and Indicators
Datacolor 245™ User's Guide
Powering Up
Use this simple procedure to start using your Datacolor 245.
1. Verify all cable connections. The AC power cord should be plugged into an
outlet. The instrument cable should be attached to the instrument and host
computer.
2. Turn on the host computer.
3. Turn on the power switch. It is the red switch on the back of the instrument.
−
When power is applied, all mechanisms are automatically reset.
−
The lights on the status panel will flash, and then the Ready light will stay
lit.
NOTE
If the instrument power is not turned on before a Datacolor program is
launched, you may receive an error message.
4. When
the
Ready
light remains on, your instrument is ready to use.
5. Launch a Datacolor program.
Instrument Calibration
The instrument should be calibrated every 8 hours to compensate for changes in the
environment. A black trap, white tile and green tile are provided with all Datacolor
instruments to complete the calibration:
•
The black trap and white tile are used each time the instrument is calibrated.
•
The green tile is used to perform an optional diagnostic test.
Содержание 245
Страница 1: ...Datacolor Datacolor 245 User s Guide ...
Страница 4: ......
Страница 8: ...4 Accessories Datacolor 245 User s Guide This page intentionally left blank ...
Страница 17: ...Datacolor 245 User s Guide Cable Installation 13 The Systems Properties dialog box displays ...
Страница 24: ...20 Controls and Indicators Datacolor 245 User s Guide This page intentionally left blank ...
Страница 28: ...24 Sample Presentation and Measurement Datacolor 245 User s Guide This page intentionally left blank ...
Страница 31: ...Datacolor 245 User s Guide Appendix 27 Appendix Datacolor 245 Optical Block Diagram ...