Shodex RI-101
6 Troubleshooting
6 Troubleshooting
When the solution to a problem cannot be found easily, a recording of the
communication between
Clarity
and the detector will significantly help
DataApex
support.
The data recording can be enabled by adding or amending the
COMMDRV.INI file in the
Clarity
installation directory (C:\CLARITY\CFG
by default). The file can be edited in any text editor (e.g. Notepad). The
following section should be edited or added:
[COM1]
echo=on
textmode=on
filename=CommDrvCOM1_%D.txt
reset=off
Note:
Instead of COM1, type the communication port used to communicate with
the
Shodex RI-101
detector. This port number is displayed when the
Det
Status
button in the
Method Setup - Acquisition
dialog is invoked.
Note:
%D
(or
%d
) in the filename parameter means that the log will be created
separately for each day. The
reset=off
parameter disables deleting the
content of the log each time the station is started during the same day.
The created *.TXT files will be of great help in the diagnosis of not
documented errors and communication issues.