4 Using the control module
Description
Description of the method (this description is not sent to or read from GC, it
serves for the information purposes only).
4.2.2 Method Setup - GC - FLOW 1(2)
Fig 7: Method Setup - GC - FLOW 1
This tab governs the settings concerning the injection port of the
Shimadzu GC- 201x
chromatograph. Up to two injection ports may be
configured and used at the same time. The tab is only present when some
inlet is configured in the
Shimadzu GC-201x Setup - Piping
dialog for a
given injector port. The fields dealing with flow of gases are available only
for models of chromatograph equipped with Advanced Flow Control
(
AFC
). These are
GC2010
,
GC2014
and
GC2014 APC/AFC
.
On/Off
Turns on/off the inlet.
Split Mode
Defines the operation mode for the injector port. The possible options are
Split
,
Splitless
and
Direct
(see
Shimadzu GC-201x
hardware manual for
more details on these injection modes). Available for
SPL
inlet only.
Control Mode
Sets the mode of operation of the dual-packed injector - this injector may
work in the
Dual
,
Single L
and
Single R
modes. When
Single L
or
Single R
is selected either on the
Flow 1
or
Flow 2
tab, commands located on
Flow
3
tab are automatically disabled from altering.
Sampling Time [min.]
Sets the sampling time in minutes for the
Splitless Sampling Mode
.