4 Using the control module
Clarity Controls
4.2 Method Setup - GC
4.2.1 Method Setup - GC - Oven/Zones
Fig 6: Method Setup - GC - Oven/Zones
Oven Max
Sets the maximal allowed temperature for the column oven. Possible
values are:
0
-
470
.
Note:
Please note that the maximum value may differ for certain GC models -
listed values for it range between
350
and
470
°C.
Equilibration Time
Time that the GC needs to reach the temperature equilibrium after the
temperature program gets to the set initial temperature. This time is used
to distribute the heat evenly in the oven.
Gradient Table
Sets the
Heat Rate
, desired temperature (
Final Temp
) and time to hold
this temperature (
Hold Time
). The
Total Time
column will be calculated
automatically. In other words, for isothermal part of the temperature
gradient, the temperature set in
Final Temp
will be kept for the interval set
in
Hold Time
. To delete a row, simply delete the value in the
Heat Rate
column on that row.
Temperature Zones
Sets the temperature and maximum temperature of the zones that have
been set in the
Shimadzu GC-201x Setup - Heater Ports
dialog (for more
details see also chapter
Shimadzu GC-201x Setup
on pg
24
).