3 Installation
Clarity Hardware
3.2 Standard cable for Clarity station
Fig 4: Cables for the
Clarity
station package
The standard
Clarity
station package includes a cable composed of
signal and starting cables for connecting the
Clarity
station to the
chromatograph.
Signal cables
Labeled
DET1
…
DET4
, the cables are supplied as standard without
connectors [bare stripped and tinned endings – red (+), white (-) and
shielding (analogue ground)].
Starting (marker) cables
Labeled
IN1
…
IN4
, ended with CINCH connector (female). The following
cables can be connected to this CINCH connector:
l
A cable terminated with free leads [red (+), shielding (digital ground)] for
direct connection to the chromatograph.
l
A cable terminated with a button for cases where a starting contact is not
available and it is necessary to perform a manual start.
Both types of cable are supplied for each starting cable.
Digital output cables
Relay contacts labeled
OUT1R
…
OUT4R
, terminated with free leads.
These are used for synchronizing autosamplers without an
AS Control
module in the
ACTIVE
Sequence or for controlling other devices.
At the converter end the cable is always terminated with a DB- 37F
connector (37-pin female connector).
- 6 -