INT7 A/D Converter
4 Using the INT7 Card
4.2 Method Setup
Parameters of the
and
dialogs depend on the type of analysis (measuring
conditions), so they can be specific for various types of analyses. This is
the reason for them to be a part of the method.
Caution:
The parameters can't be changed during analysis.
4.2.1 Method Setup - Acquisition
Fig 13: Method Setup - Acquisition (INT7)
Select Detector
Selects the detector whose parameters are set. If there is only one
detector (channel) configured on the
Clarity
Instrument, this control is not
accessible.
Range
Selects the input voltage range (in mV). The selected range corresponds
to the maximum output voltage of the detector that the A/D converter can
still process. In case the detector unit conversion is set in
dialog, the range will be shown in detector units according to the
conversion, while the corresponding original range in mV will be shown
next to the field.
Available ranges for the
INT7
card are
156
,
1 250
and
10 000 mV
.
Sample Rate
Specifies the sampling rate from among
10
,
25.04
,
50.08
, or
100.16
samples per second.
- 15 -