Troubleshooting
129
Device failure
error reported
The DT9857 or DT9857E module cannot
communicate with the Microsoft bus driver
or a problem with the bus driver exists.
Check your cabling and wiring and tighten any
loose connections; see the instructions in
.
The DT9857 or DT9857E module was
removed while an operation was being
performed.
Ensure that your module is properly connected;
see the instructions in
Data appears to
be invalid
An open connection exists.
Check your wiring and fix any open
connections; see the instructions in
.
A transducer is not connected to the
channel being read.
Check the transducer connections; see the
instructions in
The module is set up for differential inputs
while the transducers are wired as
single-ended inputs or vice versa.
Check your wiring and ensure that what you
specify in software matches your hardware
configuration; see the instructions in
.
The module is out of calibration.
DT9857 and DT9857E modules are calibrated
at the factory. If you want to readjust the
calibration of the analog input or analog output
circuitry, refer to
.
USB is not
recognized
Your operating system does not have the
appropriate Service Pack installed.
Ensure that you load the appropriate Windows
Service Pack. If you are unsure of whether you
are using USB 3.0, 2.0, or USB 1.1, run the
Open Layers Control Panel applet, described in
.
Standby mode is enabled on your PC.
For some PCs, you may need to disable
standby mode on your system for proper USB
operation. Consult Microsoft for more
information.
Table 23: Troubleshooting Problems (cont.)
Symptom
Possible Cause
Possible Solution
Содержание DT9857
Страница 1: ...DT9857 and DT9857E UM 25656 N User s Manual Title Page ...
Страница 4: ......
Страница 10: ...Contents 10 ...
Страница 14: ...About this Manual 14 ...
Страница 15: ...15 1 Overview Hardware Features 16 Supported Software 19 Supported Accessories 22 Getting Started Procedure 24 ...
Страница 25: ...Part 1 Getting Started ...
Страница 26: ......
Страница 40: ...Chapter 2 40 ...
Страница 61: ...61 4 Verifying the Operation of a Module Select the Device 63 Acquire Data from an Analog Output Channel 65 ...
Страница 64: ...Chapter 4 64 ...
Страница 71: ...Part 2 Using Your Module ...
Страница 72: ......
Страница 75: ...Principles of Operation 75 Figure 32 Block Diagram of the DT9857E Module ...
Страница 126: ...Chapter 6 126 ...
Страница 127: ...127 7 Troubleshooting General Checklist 128 Technical Support 130 If Your Module Needs Factory Service 131 ...
Страница 132: ...Chapter 7 132 ...
Страница 140: ...Chapter 8 140 ...
Страница 180: ...Appendix B 180 ...
Страница 181: ...181 C Register Level Programming Writing to the EEPROM Register 182 ...
Страница 184: ...Appendix C 184 ...
Страница 192: ...Index 192 ...